| Fan: |
|
| Flow: |
Accuracy ±5%, ±3% with Auto Correct in Retrotec software |
| Maximum Flow in Free Air |
6,300 CFM | 2,973 Litres/s | 10,705m3/h |
| Maximum Flow at 50 Pascal test pressure: |
5,600 CFM | 2,643 Litres/s | 9,514m3/h |
| Minimum Flow on Range Configuration C1 at 10 Pa |
38 CFM | 18 Litres/s | 65 m3/h |
| Dimensions: |
|
| Height: |
26" (66cm) |
| Depth: |
9.5" (24.1cm) |
| Inlet Diameter: |
22" (56 cm) |
| Weight: |
34lbs (15.4kg) |
| Motor: |
0.75hp, 9.8 amps at 120V, 4.5 amps at 240V |
| Calibration meets: |
Canadian CGSB-149.1-2002 & Super E, USA ASTM E779 – European EN 13892 & CEN, UK TM-23 Energy Standards, Rain Penetration Standards. NFPA 1001 - 01, ISO 14520 and UBC-905 fire standards |
| Door Panel: |
|
| Panel Case: |
31" x 30" x 7" (77cm x 76cm x 18cm) |
| Panel Height (min/max): |
76.5" to 87" (194cm to 221cm) |
| Panel Width (min/max): |
28.5" to 41.5" (72.4cm to 105.4cm) |
| Panel Weight: |
21.5lb (9.8kg) |
| DM-2 Digital Pressure Gauge: |
|
| Pressure Chanels: |
2 independent pressure channels (A & B) |
| Pressure Range: |
-1250 Pa to +1250 Pa (-5" to +5" H2O |
| Resolution: |
0.01 Pa (0.0001" H2O, 0.001 lb/ft2) |
| Accuracy: |
1% of pressure reading or 0.15 Pa (whichever is greater) |
| Weight: |
14oz (397g) |
| Dimensions: |
8.75" x 4.25" x 3" (22.2cm x 10.8cm x 7.6cm) |
| LCD Display: |
2.4" x 1.4" (6cm x 3.6cm) |
| Operating Temperature Range: |
32°F to 130°F (0°C to 53°C) |
| Display Backlight: |
Press [ON] to activate |
| Power: |
4 x NiMH rechargable batteries and charger included |
| Auto Off: |
4 hours |
| Temperature: |
Can be displayed with Retrotec’s temperature probe (optional) |
| Calibration meets: |
Canadian CGSB-149.1-2002 & Super E, USA ASTM E779 – European EN 13892 & CEN, UK TM-23 Energy Standards, Rain Penetration Standards. NFPA 1001 - 01, ISO 14520 and UBC-905 fire standards |
| Supported Units of Measurement: |
|
| Channel A: |
Pressure in Pascals, inches H2O, lb/ft2 |
| Channel B: |
Pressure in Pascals, inches H2O, lb/ft2
OR Flow in cfm, l/s, m3/s, m3/h
OR Flow @ (any pressure) calculates ELA at ANY desired pressure configured in the Setup menu
OR Leakage Area – EqLA (Canadian) and/or EfLA (US) in cm2, in2, ft2
OR Leakage Area @ (any pressure) calculates ELA at ANY desired pressure configured in Setup menu
OR Air-changes per hour according to volume entered in keypad
OR Velocity in m/s, km/h, ft/s, ft/min, mph
OR Velocity-Flow in cfm, l/s, m3/s, m3/h according to cross-sectional area entered on keypad |
| Functionality: |
|
| Range/Config: |
Cycles through flow ranges of the blower door or duct tester being used |
| Setup: |
Permits changing default units, displayed outputs, devices and reference pressures |
| Baseline: |
Measures and removes background pressures on Channel A for subequent testing |
| Time Avg: |
Advanced box-car averaging displays continuous updates of the average pressure. No more waiting for your next average pressure measurement! Dramatically reduces the affect of wind and other pressure fluctuations |
| Auto Zero: |
On start-up and every 8 seconds on both channels. May be turned off |
| Device: |
Selects the flow characteristics of virtually any blower door, duct-tester or a pilot tube |
| @: |
Turns on/off extrapolation to configured pressures |
| Volume/Area: |
Enter volume and/or area values to calculate air-change and flow/unit area results. |
| Temperature: |
Displayed with Retrotec’s temperature probe (optional) |